Email Medtest DX Medtest Headquarters Medtest Facebook Medtest Linkedin

Click to Play


MedTest 800 Ensemble


(Chemistry)


What makes the BA-800M analyzer different from the rest? 

The BA-800M analyzer is truly unique compared to other analyzers in its class.  What makes it stand out?

    1. Largest Walk-Away Capacity - The BA-800M analyzer offers the Largest Number of Sample Positions:
      • 440 on the BA-800M
      • 580 on the BA-800M2
    2. Most economical reagent consumption - The BA-800M analyzer has the Least Amount of Reagent Consumption - Minimum Reaction Volume is 100 µL.
    3. Scalable - Add a 2nd BA-800M system, creating a BA-800M2, to increase sample and through-put capacity to meet growing testing volumes. The BA-800M2 has 2 analytical processers and one Sample Delivery Module.
    4. Specialty Test Menu Capable - The BA-800M analyzer accommodates up to 4-reagent testing formats (example - Vitamin D assay

Visit the "Menu" tab below to see our list of chemistry reagents. For more detailed information about our chemistry test menu, please contact us at marketing@medtest.com

  • Analyzer
  • Menu
  • Consumables
  • Service & Support

Ordering Information

Catalog:

Description:

80-BA800M-ISE

BA-800M Analyzer with ISE

(One Year Warranty Included)

Onsite Installation and Initial Training Included

Benefits

CLINICAL VALUE

·      Large Test Menu – Up to 4 reagent formats

·      High Through-Put – Generate patient results faster

·      STAT Function – Fast Lane for STAT Rack on SDM

·      Reflexive Testing – Programmable to minimize wait time for results

·      Serum Indices – Alerts to integrity of sample from hemolysis, icterus, or lipemia

·      Sample Clot Detection and Reagent Bubble Detection

·      Carryover Prevention = Confidence in Results

OPERATIONAL VALUE

·      Scalable – Add a 2nd 800M system to increase sample and through-put capacity

·      Sample Clot Detection and Reagent Bubble Detection

·      Probe Collision Protection (vertical & horizontal) with auto-recovery function minimizes operator time

·      Largest Number of Sample Positions – Operator doesn’t need to feed samples as often

·      Built-in backup should the SDM or sample carousel malfunction

·      Auto-Dilute and Pre-Dilute Functions are programmable saving operator time

·      Five color-coded racks to facilitate STAT, Routine, Calibrator, Control, and Re-Run samples

·      8-grade, double probe wash station = Thorough cuvette cleaning and drying

·      Easy-to-use touch-button software

·      Step-by-step instruction guide built into software

FINANCIAL VALUE

·      Minimum Reaction Volume of 100 µL means less reagents used per test means lower cost per test and more tests per kit

·      Durable Glass Cuvettes – Replace cuvettes less often

·      Direct Solid Heating of Reaction Carousel – No consumable costs for heating oil’s and no maintenance needed to unit

·      Special Design of Reagent Bottles allows for less dead volume waste

·      Prolonged life of Tungsten Lamp – Voltage is adjusted based on lamp status, so there is no need to replace as often

Specifications

System Function

Fully automated, discrete, random access, STAT, urine and homogeneous immunoassays; STAT sample priority

Throughput:

800 photometric tests/hour (1,200 with ISE)

Measuring Principles:

Absorbance Photometry, Turbidimetry

Methodology:

End-Point, Fixed-Time, Kinetic

Single/Dual/Triple/Quadruple Reagent Chemistries

Monochromatic/Bichromatic

Programming:

User defined profiles and calculation

 

 

ISE Module

Principle:

Indirect K+, Na+, Cl-

Sample Handling

Total Sample Capacity:

440 (140 positions on the Sample Tray and 300 positions on the Sample Delivery Module)

Sample Tray:

25 cooled positions for calibrators and controls

Sample Volume:

1.5~50.1 μl, step at 0.1 μl (22 μl for ISE)

Sample Probe:

Liquid Level Detection, Clot Detection, and Collision Protection

Reagent Handling

Reagent Tray:

120 positions

Reagent Volume:

15~300 μl, step at 0.5 μl

Reagent Probe:

Liquid Level Detection, Bubble Detection, and Collision Protection

Reaction System

Reaction Volume:

100~360 μl

Operating Temperature:

37° C

Temperature Fluctuation:

± 0.1° C

Optical System

Light Source:

Halogen-Tungsten Lamp

Photometer:

Reversed Optics, Grating Photometry

Wavelength:

340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm

Absorbance Range:

0~3.4 Abs (10mm conversion)

Barcode and Data Transmission

Codabar, ITF (Interleaved Two of Five), Code128, Code39, UPC/EAN, Code93

Bi-Directional LIS Interface Transmission

Control and Calibration

Calibration Mode:

Linear (one-point, two-point and multipoint)

Logit-Log 4P, Logit-Log 5P, Spline, Exponential

Polynomial, Parabola

Control Rules:

Westgard Multi-Rule, Levy-Jennings, Twin Plot

Operation Unit

Operating System:

Windows® XP/Professional / Home SP2 or above, Windows® 7, Windows® 8

Working Conditions

Dimension:

40in x 47in x 91in (D x H x L)

Weight:

1,430 lbs